Calibration System

Accelerometer Calibration System SCS-ACV

  • Back-to-Back Calibration Installing;
  • According to ISO16063-21;
  • Support PR, IEPE, DIR Sensor Types;
  • Measure Offset, Bias Voltage, Bridge Resistance;
  • Frequency Range 20Hz~5000Hz;
  • Vibration Max. Moving 8mm;
  • Transverse Moving≤5%;
  • Sampling Frequency≥500kHz, 16bit A/D;
  • Force Rating 45N Sine Peak;
  • Peak≥20g@150g Payload; Max. Payload 500g;
  • Expanded Uncertainty (k=2): 1.0% (80Hz~160Hz); 1.5% (20Hz~2500Hz); 2.5% (2500Hz~5000Hz).
Accelerometer Calibration System SCS-ACV, according to ISO16063-21, is very suitable to middle-high frequency range (20Hz~10kHz). System include a vibration exciter BK4809, which is very suitable to small object. And more force rating type BK4808 is optional if needed. The standard accelerometer Endevco 2270 is good at calibration application (Resonant Frequency≥35kHz). System control and report generation are by software automatically.
 

Technical Specification (22℃±2℃, 30%RH~75%RH):
Name Unit Value
Frequency Range Hz 20~5000
Force Rating N 45
Max. Moving mm 8
Calibration Result Sensitivity, Deviation, Offset…
PR Excitation V 5 or 10
PR Signal mV ±1000
IEPE Constant Current mA 4
IEPE Max. Voltage V 21
IEPE Signal V ±10
DIR Signal V ±10
Aided Power Output V 24,±12
Vibration Exciter B&K 4809,Max. Acc 75g
Max. Velocity 1.65m/s
Max. Driver Current 5A RMS
Operation Temp. 5℃~50℃
Power Amplifier B&K 2718,Max. 75VA
Output Max. Current 5A RMS
10Hz~20kHz Response Error <±0.5dB

Standard Accelerometer Endevco 2270:
Name Unit Value
Sensitivity pC/g 2.2
Response (-3dB) Hz 2~20000
Transverse Sensitivity <3%
Amplitude Error per 1000g <0.1%
Sensitivity Stability <±0.2% / year
 
Data Acquisition Device
Name Unit Value
Signal Input Range V ±10
Sampling Frequency kHz 500
Resolution 16bit; Gain Error<200ppm
Input Impedance >100
 
System Diagram:


Accelerometer Calibration System SCS-ACV