Calibration System

Accelerometer Calibration System SCS-ACV

  • Back-to-Back Calibration Installing;
  • According to ISO16063-21;
  • Support PR, IEPE, DIR Sensor Types;
  • Measure Offset, Bias Voltage, Bridge Resistance;
  • Frequency Range 20Hz~5000Hz;
  • Vibration Max. Moving 12.7mm;
  • Transverse Moving≤5%;
  • Sampling Frequency≥500kHz, 20bit A/D;
  • Force Rating 112N Sine Peak;
  • Peak≥20g@400grams Payload;
  • Expanded Uncertainty (k=2): 1.0% (80Hz~160Hz); 1.5% (20Hz~2500Hz); 2.5% (2500Hz~5000Hz).
      Accelerometer Calibration System SCS-ACV, according to ISO16063-21, is very suitable to middle-high frequency range (20Hz~5kHz). System include a vibration exciter BK4808, which is very suitable to small object. And force rating is 112N without cooling. The standard accelerometer Endevco 2270 is good at calibration application (Resonant Frequency≥35kHz). System control and report generation are by software automatically.

Accelerometer Calibration System SCS-ACV