Back-to-Back Calibration Installing;
- According to ISO16063-21;
- Support PR, IEPE, DIR Sensor Types;
- Measure Offset, Bias Voltage, Bridge Resistance;
- Frequency Range 20Hz~5000Hz;
- Vibration Max. Moving 8mm;
- Transverse Moving≤5%;
- Sampling Frequency≥500kHz, 16bit A/D;
- Force Rating 45N Sine Peak;
- Peak≥20g@150g Payload; Max. Payload 500g;
- Expanded Uncertainty (k=2): 1.0% (80Hz~160Hz); 1.5% (20Hz~2500Hz); 2.5% (2500Hz~5000Hz).
Accelerometer Calibration System SCS-ACV, according to ISO16063-21, is very suitable to middle-high frequency range (20Hz~10kHz). System include a vibration exciter BK4809, which is very suitable to small object. And more force rating type BK4808 is optional if needed. The standard accelerometer Endevco 2270 is good at calibration application (Resonant Frequency≥35kHz). System control and report generation are by software automatically.