- Back-to-Back Calibration Installing;
- According to ISO16063-22;
- Support PR, IEPE, DIR Sensor Types;
- Measure Offset, Bias Voltage, Bridge Resistance;
- Calibration Range 25g~2000g;
- Shock Pulse Width 0.1ms~3ms;
- Transverse Moving≤5%;
- Sampling Frequency 500kHz, 16bit A/D;
- Support Continues Shock by Compressed Air;
- Max. Payload≥50g;
- Expanded Uncertainty (k=2): 1.8%.
The accelerometer shock calibration system SCS-ACS adopts the principle of piston pneumatic shock to generate physical acceleration, and is designed with reference to the ISO16063-22 standard. Under the automatic control of the software, by controlling the shock pressure and the valve switch, the half-sine acceleration peak value of 25g~2000g is obtained. Endevco 2270, as the reference standard acceleration sensor, has long term stability with high resonant frequency. The system is automatically controlled by software, the calibration report is automatically generated, and the report template can be customized according to requirements.
Technical Specification (22℃±2℃, 30%RH~75%RH):Name | Unit | Value |
Calibration Range | g | 25~2000 |
Shock Pulse | ms | 3.0~0.1 |
Payload | grams | ≥50 |
Calibration Result | Sensitivity, Non-Linearity, Offset, Bias Voltage… | |
PR Excitation | V | 5 or 10 |
PR Signal | mV | ±1000 |
IEPE Current | mA | 4 |
IEPE Max. Voltage | V | 21 |
IEPE Signal | V | ±10 |
DIR Signal | V | ±10 |
Aided Power Output | V | 24,±12 |
Power | VAC | 100~250 |
Compressed Air | bar | ≥5 |
Standard Accelerometer |
Endevco 2270 Response(-3dB) 2Hz~20kHz Sensitivity error≤0.1% per 1000g |
System Diagram:
Procedure:
1. Mounting the Unit of Under Test (UUT) and Connecting with the correct channel of DAS;
2. Setup sensor information, Excitation, Calibration Range, etc. and Warm up;
3. Click Run by Software, and wait until all shocks had been done and report generated;
4. Remove UUT for the Next one or Shunt down.
Procedure:
1. Mounting the Unit of Under Test (UUT) and Connecting with the correct channel of DAS;
2. Setup sensor information, Excitation, Calibration Range, etc. and Warm up;
3. Click Run by Software, and wait until all shocks had been done and report generated;
4. Remove UUT for the Next one or Shunt down.
Accelerometer Calibration System SCS-ACS |